HICO 555
HICO-VARIOTHERM 555 is the allrounder among the hypo-/hyperthermia systems and appreciated for its reliable warming and cooling of patients in intensive care units and operating rooms. Featuring a water temperature range of 5 to 40 °C, this system covers all areas of application.
- Easy and quick start-up of the non-invasive temperature control system
- Optimum temperature transfer via reusable HICO water pads
- Connections for two HICO water pads
- Optional use of Soft-Temp® single-use pads
- Nursing staff has free access to the patients since they are positioned on the water pads.
Technical Data
REF
|
600001 (230 V) / 600021 (115 V)
|
Nominal voltage
|
230 VAC 50 Hz / 115 VAC 60 Hz
|
Power input
|
440 W
|
Current consumption
|
approx. 1.9 A (230 V) / 3.8 A (115 V)
|
Nominal temperature range
|
5 °C – 40 °C
|
Safety shutdown
|
approx. 4 °C and 41.0 °C (independent safety shutdown, resulting in a max. pad surface temperature of < 41 °C)
|
Measuring range
|
approx. 0 °C – 45 °C
|
Measurement error 1
|
< +0.2 °C (display – water temperature)
|
Measurement error 2
|
< +0.7 °C (display – pad surface temperature)
|
Correction value
|
0.5 °C (water temperature – temperature display)
|
Pump capacity
|
max. 7 l/min, max. 0.36 bar
|
Warm-up time
|
approx. 5 – 10 min (20 - 37 °C without water pad attached)
|
Cool-down time
|
approx. 5 – 10 min (20 - 10 °C without water pad attached)
|
Fuse rating
|
2 x T 3,15 A; L 250 V (230 V) /
2 x T 6,3 A; L 250 V (115 V)
|
Class / type of protection
(IEC 60601-1)
|
I / BF (defibrillation-proof)
|
IP protection class (IEC 60529)
|
IP X1 (drip-proof)
|
Risk class (93/42/EEC)
|
II b
|
Ambient temperature
|
10 – 40 °C
|
Relative humidity
|
approx. 30 – 70 %
|
Transport / storage temperature
|
3 – 60 °C
|
Air pressure
|
700 – 1060 mbar
|
Tank volume
|
approx. 1.0/2.0 l (min./max.)
|
Permissible height difference
|
max. 1 m (between unit and water pad)
|
Dimensions W x H x D
|
approx. 20 x 60 x 51 cm
|
Packaging dimensions
|
295 x 795 x 590 mm
|
Weight
|
approx. 25 kg (empty water tank)
|
Noise emission
|
approx. 45 dB(A)(1 m)
|
Alarm level
|
> 55 dB(A)(3m)
|
Tank volume heat exchanger
|
1.5 l
|
Coolant
|
water-glycol mixture
|
UMDNS code
|
12-075
|
For Details Please Contact Us At
07-3310512 / 07-33138423
No comments:
Post a Comment